User`s guide

M3000A/M3046A/M3015A/M3016A Service Guide
2-22 Installing the Instrument
Site Preparation Guidelines
Although the Instrument (Monitor and Measurement Server or, where present, Measurement
Server Extension) is primarily a portable Monitor, some site preparation must be done before
using it in certain locations. Additionally, good operating practices are also recommended.
Check that the operating environment is reasonably free from dust, vibration, corrosive or
explosive gases and is within specified temperature and humidity ranges.
Position the Instrument so that the display is clearly visible and the controls are easy to
reach.
The M3046A 0867 be grounded during operation. If a three-wire receptacle is not availa-
ble then a hospital electrician must be consulted to ensure that proper grounding is available
on installation. 1(9(5 attempt to use a three-wire to a two-wire adapter with the
M3046A. The Instrument must be grounded whenever it is connected to the mains power
using a 3-pin plug.
Installing the Table Mount (M3080A #A10)
The Table Mount is intended for any flat, level surface. To mount the plate securely, you need
4 screws. These are not provided because the type of surface dictates the type of screw.
6WHS Unpack the Table Mount and check all parts are available.
6WHS Select 4 screws (appropriate to surface material).
6WHS Insert the 4 screws through the screw holes provided in the plate.
6WHS Tighten the screws.
The Monitor snap-fits onto the plate. To release the Monitor, press the button on the front of
the plate.