Installing the Instrument 2-15
Installing an Additional Display
An additional display must be installed by an Agilent Technologies service engineer or
authorized Service Representative. By the addition of a display, the M3/M4 Monitor becomes
a system and must be tested as such after installation. The required tests are described in
Chapter 4, Testing and Inspecting.
There are two displays available from Agilent Technologies which have been tested for use
with the Monitor. They can be ordered under the following Monitor option numbers:
0$+15" Color Display (IEC 601-1 compliant, for use in the patient vicinity).
0$+21" Color Display. An isolation transformer must be used with this display
and the HP 1389A isolation transformer is included in this option (the combination complies
with IEC 601-1-1, for use outside the patient vicinity).
Other displays which meet the specifications for the VGA interface may also be used. These
displays must either:
Comply with IEC 601-1, when used in the patient vicinity, or
Be used with an isolation transformer (e.g. HP M1389A) outside the patient vicinity. The
combination must comply with IEC 601-1-1.
,QVWDOODWLRQ
A Video Out connector is provided on the rear of the Monitor, as shown below.
LAN / Software Update
Nurse Call Relay
Connector (
36V)
Connector (≤ 5V)
Video out Connector
for an additional display
(
3.3V)
Screw connector
for protective
earth