User`s guide
M3000A/M3046A/M3015A/M3016A Service Guide
Installing the Instrument 2-13
6WHS Press and hold the mount latch toward the clamp screw.
Rotate the Measurement Server and mount until you get it to the position you want.
6WHS Release the mount latch, and make sure it is clicked into one of the four slots on the
back of the mount.
Connecting to the Nurse Call Relay
The nurse call relay is a 3.5mm, mono phone jack socket. It is completely isolated from the
rest of the circuitry. Under normal conditions, the tip and sleeve are “open” (not shorted
together). When an alarm is indicated, the tip and sleeve are shorted together by a relay.
:$51,1*
'RQRWUHO\H[FOXVLYHO\RQWKH1XUVH&DOO5HOD\IRUWKHQRWLILFDWLRQRIDODUPFRQGLWLRQV
7KHUHOD\RXWSXWFDQQRWEHFKHFNHGE\WKH0RQLWRUDQGWKH0RQLWRUFDQQRWQRWLI\WKH
XVHURIDQ\IDLOXUHRIWKHUHOD\
See the specifications for the Nurse Call Relay in the Monitor and Measurement Server
Specifications chapter of the User’s Guide, and the documentation for the device you are
connecting.
Installation of Wireless Infrastructure
See Agilent LAN Database Server Installation and Service Manual M3154-90000.
Mount