User`s guide
M3000A/M3046A/M3015A/M3016A Service Guide
Introduction to the Instrument 1-53
0DLQVWUHDP&2
:DYH
The two calibration marks are located at 10% and 60% of wave channel height. The lower
calibration mark is labeled 0 mmHg (0.0 kPa) on all scales. The upper one is adjustable from
20 mmHg (2kPa) to 100 mmHg (13 kPa) in steps of 10 mmHg (1kPa).
0HDVXUHPHQW0RGH
In the measurement mode, the Extension continuously measures CO
2
concentration and
calculates medical parameters.
• 1
2&RUUHFWLRQ—This can be turned on or off. If N
2
O correction is off, only oxygen cor-
rection to CO
2
is made. In this case, it is assumed that the gas mixture consists mainly of O
2
and N
2
(respiratory intensive care unit conditions).
If N
2
O correction is on, oxygen and N
2
O correction to the CO
2
value is made. In this case,
it is assumed that the gas mixture consists mainly of O
2
and N
2
O (operating room condi-
tions).
• 2
&RUUHFWLRQ— There is a fixed correction of 45% O
2
.
• +XPLGLW\&RUUHFWLRQ—This correction is selectable between Body Temperature Pressure
Saturated (BTPS) and Standard Pressure Temperature Dry (STPD). The Extension meas-
ures BTPS and uses this correction formula:
where P
H20
= 47mmHg
• 0D[+ROG—There are three possible selections:
• 2II—The EtCO
2
numerics display the breath-to-breath value.
• VHFRQGV—The EtCO
2
numerics display the highest value within a moving window
over a 10-second period.
• VHFRQGV—The EtCO
2
numerics display the highest value within a moving window
over a 20-second period.
6DIHW\
To ensure the safety of the patient, the patient-applied part is isolated from ground by opto-
couplers and a transformer. The circuit is also encapsulated in plastic.
Calibration Marks
Wave Range
&2
P
STPD
= P
BTPS .
P
abs
P
abs
–P
H2O