User`s guide

M3000A/M3046A/M3015A/M3016A Service Guide
Introduction to the Instrument 1-27
1%30RGHV
The measurement offers adult, pediatric, and neonatal modes.
The following table lists the cuff inflation limits for each mode.
The following table lists the measurement ranges for each mode.
6DIHW\
The following table lists the maximum limits that ensure patient safety.
If any one of these safety limits is violated, an INOP is generated and the valve opens.
0RGH )LUVW,QIODWLRQ
6XEVHTXHQW,QIODWLRQV
$ERYH6\VWROLF3UHVVXUH
6WDW0RGH
Adult 165 25 15
Pediatric 130 20 15
Neonatal 100 15 15
0RGH 6\VWROLF 'LDVWROLF 0HDQ
Adult 30 - 270 10 - 245 20 - 255
Pediatric 30 - 180 10 - 150 20 - 160
Neonatal 30 - 130 10 - 100 20 - 120
0RGH
0D[LPXP
0HDVXUHPHQW7LPH
0D[LPXP7LPH
3UHVVXUH
2YHUSUHVVXUH
0D[LPXP
Adult 180 seconds 180 seconds for
pressure > 15
mmHg
300 mmHg for > 2
seconds
Pediatric 180 seconds 180 seconds for
pressure > 15
mmHg
300 mmHg for > 2
seconds
Neonatal 90 seconds 90 seconds for pres-
sure > 5 mmHg
150 mmHg for > 2
seconds