User`s guide

M3000A/M3046A/M3015A/M3016A Service Guide
6-20 Repairing the Instrument
5HILWWLQJWKH,QIUDUHG/DPS
To refit the Infrared Lamp, do the following:
6WHS Fit the spring over the infrared lamp.
6WHS Making sure the red lamp-removal-band lies flat in the bottom of the assembly, and
holding the spring onto the end of the Infrared Lamp with a finger, place the lamp
into the assembly. Make sure that the contact is oriented correctly for contact to
take place with the red band around the Infrared Lamp and that the entire assembly
(housing, lamp and spring) is seated firmly.
6WHS Replace the Infrared Lamp cover.
6WHS Being careful not to touch anything else in the Extension, use the thin-bladed
screwdriver or your fingertips to close the locking clamp holding the Infrared Lamp
assembly closed.
127( Make sure that the movement of the locking clamp does not pinch any
adjacent tubing.
127( After replacing the Infrared Lamp, reset the value displayed on the Reset
IRSourceTime selection to zero (Service Mode>CO
2
Setup).