6-18 Repairing the Instrument
5HPRYLQJWKH3XPS
To remove the Pump, do the following:
6WHS Locate the Pump in the Extension.
6WHS Being careful not to touch anything else in the Extension, unscrew the screw hold-
ing the pump bracket in position. Lift the top part of the bracket away and lift out
the pump.
6WHS Gently disconnect the flow tubing attached to the Extension from the Pump.
127( Be sure to note which tube attaches to the inlet and which tube attaches to
the outlet.
6WHS Gently disconnect the power lead which attaches the Pump to the Extension.
6WHS Remove the Pump.
127( After replacing the Pump, reset the value displayed on the Reset PumpOpTime
selection to zero (Service Mode>CO
2
Setup). When the PumpOpTime has been reset an
INOP will be generated: “CO
2
OCCLUSION”. To clear this INOP you must perform a flow
check and store the flow in Service Mode (select “Store Flow”)