User`s guide
M3000A/M3046A/M3015A/M3016A Service Guide
6-16 Repairing the Instrument
5HPRYLQJWKH&2
6FUXEEHU
To remove the CO
2
Scrubber, do the following:
6WHS Locate the CO
2
Scrubber in the Extension.
6WHS Being careful not to touch anything else in the Extension, use tweezers to pull the
body of the CO
2
Scrubber out of the bracket
.
6WHS Holding the body of the CO
2
Scrubber with your fingers, carefully disconnect the
Extension intake tube from the scrubber end and remove the CO
2
Scrubber from
the Extension.
6WHS Dispose of the CO
2
Scrubber according to local legal requirements for low volume
chemical waste.
5HPRYLQJWKH,QIUDUHG/DPS
To remove the Infrared Lamp, do the following:
6WHS Locate the Infrared Lamp assembly in the Extension.
6WHS Being careful not to touch anything else in the Extension, use a thin-bladed screw-
driver to pry open the locking clamp holding the Infrared Lamp assembly closed.
127( Make sure that the movement of the locking clamp does not pinch any
adjacent tubing.
6WHS Remove the Infrared Lamp cover