Refit Procedures for the Monitor
5HILWWLQJWKH6\VWHP%RDUG
To refit the System Board, complete the steps for removing the System Board in reverse.
When refitting the System Board please observe the following:
Be careful not to strike the metal chassis with parts.
Make sure the speaker cable is not trapped when the System Board is inserted.
Wait until the System Board is in place before positioning the screws, otherwise they may
fall out when you do the next step.
Slightly raise the end of the System Board and push down and forward to engage the retain-
ing clips on the chassis.
Make sure the bottom edge of the System Board sits correctly in the clips to give a good
electrical contact.
5HILWWLQJWKH/&'$VVHPEO\DQG%DFNOLJKW7XEHV
To refit the LCD Assembly and backlight tubes, complete the steps for removing the LCD
Assembly and backlight tubes in reverse.
127( When you push the Display Assembly back into the chassis, make sure that the 6
clips around the cushion are still in place and are not bent.
5HILWWLQJWKH&RQQHFWRU%RDUG
To refit the Connector Board, complete the steps for removing the Connector Board in
reverse.
5HILWWLQJWKH6SHDNHU
To refit the Speaker, complete the steps for removing the Speaker in reverse.
5HILWWLQJWKH3RZHU6XSSO\
To refit the Power Supply, complete the steps for removing the Power Supply in reverse.
127(When replacing the Power Supply, make sure it slides into the guide rails provided in
the chassis. Be sure to refit the anti-vibration screws.
System Board
Stop