User`s guide

M3000A/M3046A/M3015A/M3016A Service Guide
6-6 Repairing the Instrument
6WHS Remove the Power Supply as described previously in this chapter.
6WHS Remove the chassis as described previously in this chapter.
6WHS Remove the screw which secures the connector board, then pull out the Connector
Board.
6WHS Use a small screwdriver to disconnect the Display Assembly cable.
6WHS Lay the chassis face down on a clean, scratch-free surface.
6WHS Disconnect the loudspeaker cable (1).
)RU0RQLWRUVZLWK:LUHOHVV/$1DVVHPEO\Disconnect the ventilator wires (2)
from the system board. Unclip the fan assembly from the metal sheet (3) and pull
forwards to remove.
6WHS Remove the 4 System Board screws (4).
1
1
2
3
1
4
4