M3000A/M3046A/M3015A/M3016A Service Guide
6-2 Repairing the Instrument
Warnings, Cautions and Safety Precautions
Do 127 disassemble the product past the point described in these procedures.
+LJK9ROWDJH:DUQLQJVoltages dangerous to life are present in the Instrument when it is
connected to the mains Power Supply. Do not perform any disassembly procedures other
than Server and Extension removal with power applied to the Instrument. Failure to adhere
to this warning could cause serious injury or death.
5HVLGXDO+LJK9ROWDJHThere may be hazardous voltages stored in capacitors on the
Power Supply board even when the Instrument is not connected to the mains Power Supply.
High voltages are present on the Power Supply board for up to 20 seconds after disconnect-
ing the mains power cord so always disconnect the Power Supply cord and wait at least 20
seconds before removing or refitting the Power Supply board.
%DWWHU\:DUQLQJNever insert a battery without the Power Supply board being present.
You may blow the battery fuse on the Connector Board.
3RZHU&RUG:DUQLQJDo not connect a mains power cord to a Power Supply that has
been removed from the Monitor.
Disassembly for the Monitor
5HPRYLQJWKH%DWWHU\
To remove the battery, do the following:
6WHS Slide the battery compartment door toward the rear of the Monitor, and open it
down.
6WHS Locate the battery extractor tape and pull it to disengage the battery socket.
6WHS Pull the battery out.
5HPRYLQJWKH3RZHU6XSSO\
5HDGWKHZDUQLQJVDWWKHEHJLQQLQJRIWKLVFKDSWHU
You must remove the Power Supply to access the fuse.
To remove the Power Supply, do the following:
6WHS Remove the Measurement Server and, where present, the Measurement Server
Extension.
6WHS Remove the battery.