User`s guide

Repairing the Instrument 6-1
6 Repairing the Instrument
Objectives
To meet the goals of this chapter, you should be able to disassemble the Instrument down to
replacement part level.
The main replaceable assemblies for the Monitor are as follows:
The Power Supply.
The System Board.
The Connector Board.
The LCD Assembly.
The Display Backlight Tubes.
The removal and replacement procedures for the Monitor are easy to complete when you
follow the step-by-step procedures in this chapter.
The main replaceable assemblies for the M3015A Measurement Server Extension are as
follows:
The CO
2
Scrubber.
The Infrared Lamp.
The Pump.
The removal and replacement procedures for the M3015A Measurement Server Extension
are easy to complete when you follow the step-by-step procedures in this chapter.
The M3000A Measurement Server and the M3016A cannot be repaired and must be
exchanged if defective.
Concepts
'LVDVVHPEO\ Disassembly refers to the removal of cover, chassis and other parts in order
to access and replace parts in the M3046A Monitor and the M3015A
Measurement Server Extension. Disassembly does not refer to a complete
breakdown of the entire Instrument since the M3000A Measurement Server
and the M3016A Measurement Server Extension cannot be disassembled.
+LJK9ROWDJH This concept refers to voltages dangerous to life. High voltages can be found
in the Instrument when it is connected to the mains Power Supply and in the
capacitors on the Power Supply board even when the Instrument is not
connected to the mains Power Supply.