User`s guide

M3000A/M3046A/M3015A/M3016A Service Guide
5-38 Troubleshooting the Instrument
7URXEOHVKRRWLQJZLWK6HOI7HVW$ODUP0HVVDJHV
:KHQ<RX6ZLWFKWKH0RQLWRU2Q
0HVVDJH :KDW7R'R
Battery Low (approximately 20
minutes remaining)
Connect to AC power to charge the battery, or fit a fully charged battery within next 15
minutes.
Check Status Log This indicates a non-critical problem in the Monitor. There is a defect but the Monitor can
still be used.
The Monitor should be investigated technically, however, as soon as possible.
ECG EQUIP MALF Measurement defective. Exchange the Measurement Server.
The Measurement Server should be investigated technically as soon as possible.
NBP EQUIP MALF Measurement defective. Exchange the Measurement Server.
The Measurement Server should be investigated technically as soon as possible.
Pressure Zero & Check Cal Pressure must be zeroed, or calibration required. Zero the Pressure, or check the calibration.
If unsuccessful, exchange the Measurement Server.
PRESS EQUIP MALF Measurement defective. Exchange the Measurement Server or Measurement Server Exten-
sion.
The Measurement Server or Measurement Server Extension should be investigated techni-
cally as soon as possible.
RESP EQUIP MALF Measurement defective. Exchange the Measurement Server.
The Measurement Server should be investigated technically as soon as possible.
SpO
2
EQUIP MALF Measurement defective. Exchange the Measurement Server.
The Measurement Server should be investigated technically as soon as possible.
TEMP EQUIP MALF Measurement defective. Exchange the Measurement Server or Measurement Server Exten-
sion.
The Measurement Server or Measurement Server Extension should be investigated techni-
cally as soon as possible.
CO
2
EQUIP MALF Measurement defective. Exchange the Measurement Server Extension or (for mainstream
CO
2
) the transducer.
BAD SERVER LINK 1) An M3000A Measurement Server with revision B software is connected to an M3046A
Monitor with revision A software. This combination does not allow monitoring.
OR
2) This combination of Monitor, Measurement server and cable/link bar cannot be used
BAD SERVER LINK
plus
Measurement Server Revision
not supported status message
in red.
An M3000A Measurement Server with revision A software is connected to an M3046A
Monitor with revision B software. This combination does not allow monitoring.
Some measurements are not
supported by the Monitor
prompt message
A measurement extension (M3015A or M3016A) is connected to an M3046A M3 monitor.
No measurements (CO
2
, 2nd Press/Temp) from the Measurement Extension are available.
Measurement Server Configu-
ration not supported status
message
A Measurement Extension (M3015A or M3016A) is connected to an M3000A Measure-
ment Server with Option #D06 (noninvasive measurements only) and an M3046A Monitor.
No measurements (CO
2
, 2nd Press/Temp) from the Measurement Extension are available.