User`s guide

M3000A/M3046A/M3015A/M3016A Service Guide
5-36 Troubleshooting the Instrument
After 15,000 to 20,000 hours (max 20 000 hours).
If the flow cannot be adjusted to the desired value.
If the Pump makes considerably more noise than new Pumps.
After replacing the pump, perform a CO
2
calibration check as described in CO2 Gas
Measurement Calibration Check on page 4-16.
127( When replacing the pump in the M3015A Measurement Server Extension
(sidestream CO
), you should also replace the CO scrubber at the same time.
5HVHW,56RXUFH7LPH
The cumulative time the infrared lamp in the M3015A Measurement Server Extension
(sidestream CO
) has been operating is shown. When replacing the infrared lamp, this
counter must be reset. A reset of the infrared lamp time must be confirmed.
The intensity of the Infrared Lamp gradually decreases over time. As a result, periodic
replacement is necessary.
Agilent Technologies recommends replacement of the Infrared Lamp as follows:
After 15,000 to 20,000 hours ( max. 20 000 hours).
If the Instrument exhibits a noticeably larger noise on the CO
2
wave.
If the Instrument does not pass the accuracy test after calibration.
6HOI7HVW&\FOHVLQIRUPDWLRQRQO\
The number of executed Self-Test cycles since the last reboot is displayed, but cannot be
modified.
2S7LPH0RQLQIRUPDWLRQRQO\
The cumulative time that the Monitor has been operated is shown, but cannot be modified.
2S7LPH0HDV6LQIRUPDWLRQRQO\
The cumulative time the Measurement Server has been operated is shown, but cannot be
modified.
1%3&\FOHVLQIRUPDWLRQRQO\
The cumulative number of NBP cycles completed is shown, but cannot be modified.
/LQH)UHTXHQF\
When operating on mains power supply, the line frequency indication is used to switch
between 50 Hz and 60 Hz for the ECG filtering frequency.