User`s guide
M3000A/M3046A/M3015A/M3016A Service Guide
5-34 Troubleshooting the Instrument
This coldstart test executes a reboot with a more intensive system Self-Test than a normal
reboot. The coldstart test is also executed in the Server and in the Extension if either one is
connected.
Executing the coldstart test automatically returns the Instrument to Monitoring Mode.
To execute a coldstart test, do the following:
6WHS Switch the Monitor into Service Mode.
6WHS Press the Setup key.
6WHS Select Status Log.
6WHS Clear the Status Log and Confirm (for both, Monitor and Server).
6WHS Press Exit.
6WHS Select Hardware Test.
6WHS Select Coldstart.
6WHS Press Confirm.
The Monitor (and Server, if connected) perform a reboot.
The coldstart reboot is the same as described in What Happens During a Regular Boot, AC
powered, without Battery H[FHSW that the time taken from the switch off of the Alarm and
Suspend LEDs to the switch on of the backlight and display of the M3 logo is extended. The
whole reboot takes about 10 seconds.
After the system has come up again, check whether a CHECK STATUS LOG INOP is
displayed.
• If not, the test is successful.
• If yes, do the following:
• Go into the Status Log and check whether either the Monitor or the Server reports a fatal
error.
• Print out the Status Log in which the error is reported.
To ensure the fault is not a single (sporadic) fault, you should rerun the coldstart sequence
(including clearing the Status Logs).
• If a fatal error is again reported, either the System Board or the Server should be replaced.
Please include the printed Status Log with the exchanged parts.