Troubleshooting the Instrument 5-21
Message appears: New display
detected, please enter new dis-
play specs”
The display has been
exchanged and the Monitor
needs the new specifica-
tions.
Use the Upgrade Tool to
upgrade the monitor with
the latest display specifica-
tions.
7RXFK6WULSUHODWHGSUREOHPV
TouchStrip discolored, cracked
or surface is deformed.
Cleaning agents, abrasive
material, or strong solvents
were used to clean the
Instrument.
Replace the Monitor Bezel
and clean the Instrument in
the future with only isopro-
pyl alcohol. Wipe cleaning
agents off surfaces immedi-
ately.
Lack of user control of display. Cable from the System
Board to the Monitor Bezel
not connected properly.
Defective TouchStrip.
Defective System Board.
Refit cable correctly.
Check Status Log for mes-
sage: CHECK INPUT
DEVICE
Replace the System Board.
1HWZRUNUHODWHGSUREOHPV
“No Central assigned to this
bed” prompt message
Wrong monitor label or
Information Center incor-
rectly configured
Check the bed assignments
at the Information Center
and that the monitor label
assigned to this monitor has
not been changed locally
“No Central - duplicate monitor
label” prompt message
Monitor label configuration
is in conflict with another
monitor
Check that the monitor label
assigned to this monitor has
not been changed locally.
Check the assignment at the
Information Center
“Assigned Central is not availa-
ble” prompt message
The Information Center is
switched off or not accessi-
ble
“No Central - software incom-
patible” prompt message
Ask Biomed department to
check software revisions
“Central can not identify this
bed” prompt message
The location of this bed is
not clear.
Check the bed assignments
at the Information Center.
0HDVXUHPHQW6HUYHUUHODWHGSUREOHPV