User`s guide
M3000A/M3046A/M3015A/M3016A Service Guide
5-20 Troubleshooting the Instrument
“Measurement Server Configu-
ration not supported” status mes-
sage
A Measurement Extension
(M3015A or M3016A) is
connected to an M3000A
Measurement Server with
Option #D06 (noninvasive
measurements only) and an
M3046A Monitor. No meas-
urements (CO
2
, 2nd Press/
Temp) from the Measure-
ment Extension are availa-
ble.
Use extensions only with
standard Measurement
Servers (not #D06).
If this message appears
when a standard Server is
connected, try restarting the
monitor.
3RZHU6XSSO\UHODWHGSUREOHPV
See Troubleshooting the Front-Panel LEDs earlier in this chapter.
'LVSOD\DQG/&'$GDSWHU%RDUGUHODWHGSUREOHPV
Monitor screen is blank. Instrument appears to boot
as it should; alarm LEDs
switch on/off, successful
test sound, and yellow LED
on the System Board blinks.
Display and other cables not
seated correctly.
See Troubleshooting the
System Board LEDs earlier
in this chapter.
Refit cables correctly.
If this does not work,
replace the LCD Display
Assembly.
Fluid dripping or appearing
below Instrument.
LCD assembly damaged or
cracked.
Replace the LCD Display
Assembly.
Decrease in light intensity on
display. Brightness decreases
from top to bottom or bottom to
top.
One of the two backlight
tubes is worn.
Replace both backlight
tubes. Although it is unusual
for both tubes to need
replacing at the same time,
you are recommended to do
so since replacement is not
easy.
Display waves are noisy/not
showing a smooth QRS wave.
Radio frequency interfer-
ence (RFI) from other
instruments such as ultra-
sound, ESU, Defib, or
telemetry is causing noisy
waveforms.
Turn off or move away from
the Monitor the instrument
causing RFI problems.