User`s guide

M3000A/M3046A/M3015A/M3016A Service Guide
Troubleshooting the Instrument 5-11
127( Do not try to remove the video flex layer because, if snapped on correctly, remov-
ing it can easily damage the SMT solder joints.
Check that the cables of the backlight tubes are seated correctly into the backlight inverter.
Check the backlight tubes are working correctly (broken, damaged, or worn).
If the display still remains dark either the Display Assembly or the System Board is defec-
tive. It is recommended to replace the Display Assembly first.
:KDW7R'RLIWKH$ODUPDQG6XVSHQG/('V$UH1RW:RUNLQJ&RUUHFWO\
If the Monitor boots until the display is switched on but the Alarm LED and the Suspend
LED have not switched as described previously (red, red/yellow then off), suspect the
following:
Check that the Monitor Bezel cable is seated correctly.
Replace the Monitor Bezel.
Replace the System Board.
,VRODWLQJ3UREOHPVWRWKH&RUUHFW6XEDVVHPEO\
:$51,1*
+LJK9ROWDJH9ROWDJHVGDQJHURXVWROLIHDUHSUHVHQWLQWKH,QVWUXPHQWZKHQLWLV
FRQQHFWHGWRWKHPDLQVSRZHUVXSSO\'RQRWSHUIRUPDQ\GLVDVVHPEO\SURFHGXUHV
RWKHUWKDQ6HUYHUDQG([WHQVLRQUHPRYDOZLWKSRZHUDSSOLHGWRWKH,QVWUXPHQW
)DLOXUHWRDGKHUHWRWKLVZDUQLQJFRXOGFDXVHVHULRXVLQMXU\RUGHDWK
7URXEOHVKRRWLQJWKH(&*287
If no marker pulse is displayed on the Monitor (assuming the defibrillator and defibrillator
cable are in working order), check the following:
Disconnect the Server and Defib cable.
Switch the Monitor off then on again. Observe the red LED in the ECG OUT section. (The
Instrument Self-Test is performed).
If the red LED does not switch on for about 1 second at power on, replace the System
Board.
If the red LED switches on and remains on for more than 20 seconds, replace the System
Board.
Use a NQRZQJRRG link bar and connect a Defib, Defib cable and the Measurement Server
again. Check marker pulse again.
Use a NQRZQJRRG Measurement Server. Check marker pulse again.
If both of the above fail, replace the System Board.
If there is no ECG OUT signal to the Defib, repeat the above checks.