User`s guide

M3000A/M3046A/M3015A/M3016A Service Guide
Troubleshooting the Instrument 5-5
7URXEOHVKRRWLQJWKH)URQW3DQHO/('V
When the Monitor is first switched on, all the front-panel LEDs and keys light up
momentarily. The meaning of the front-panel LEDs under normal operation is given in the
following table together with a brief summary of possible defect conditions:
With the Instrument on, observe the following:
The backlight tube lights the LCD display.
No error messages are displayed. Error messages indicate either software or hardware fault
has occurred. If error messages are displayed, refer to error codes later in this chapter.
The brightness of the LCD display is adequate. If not, adjust it accordingly.
The green LED on the System Board is on (see the following section).
LED Normal Operation Defect Condition
On
Off/Standby
Green: The Monitor is on.
Off: The Monitor is in standby mode.
The LED remains off: Suspect:
1. Monitor Bezel cable connec-
tion.
2. System Board
Go to the troubleshooting sec-
tion.
AC Power Green: The Monitor is connected to the
mains power supply.
The LED remains off: Suspect:
1. Power Supply.
2. Monitor Bezel cable connec-
tion.
3. System Board.
Go to the troubleshooting sec-
tion.
Battery Green: The battery is fully charged
(>95%).
Yellow: The battery is charging.
Red and blinking: The battery is empty.
Same as for AC Power condi-
tion.