M3000A/M3046AM3015A/M3016A Service Guide
Testing the Instrument 4-25
63DUW6\VWHP(QFORVXUH/HDNDJH&XUUHQW1&QRUPDOFRQGLWLRQ
Expected test results:
Normal condition maximum leakage current x1 ≤ 100µA
Measures leakage current of exposed metal parts of Instrument under Test (IUT) and between
parts of the system within the SDWLHQWHQYLURQPHQW; normal and reversed polarity using S2.
Safety test according IEC 601-1 / UL2601-1
(*) Not present in Class 2.
(**) Can be multiple different connections to different equipment at same time.
(**)