User`s guide
M3000A/M3046AM3015A/M3016A Service Guide
4-20 Testing the Instrument
— Base impedance line 1500 Ohm.
— Delta impedance 0.5 Ohm
— Respiration rate 40 rpm.
6WHS The value should be 40 rpm +/- 2 rpm.
,QYDVLYH3UHVVXUH3HUIRUPDQFH7HVW
This test checks the performance of the invasive pressure parameter. The Invasive Pressure
performance test is required once every two years.
7RROVUHTXLUHG Patient Simulator.
6WHS Connect the Patient Simulator to the Pressure connector on the Measurement Server
or the Measurement Server Extension.
6WHS Set Patient Simulator to 0 pressure.
6WHS Make a zero calibration.
6WHS Configure the Patient Simulator as follows:
— P(static) = 200 mmHg.
6WHS Wait for the display.
6WHS The value should be 200 mmHg ± 5 mmHg. If the value is outside these tolerances,
calibrate the Measurement Server or Measurement Server Extension.
127(
If the Measurement Server was calibrated with a dedicated reusable catheter, check the
calibration together with this catheter.