M3000A/M3046AM3015A/M3016A Service Guide
Testing the Instrument 4-19
Accuracy and Performance Procedures
The following accuracy, calibration, and performance procedures are designed to be
completed to verify the accuracy and performance of the Instrument. They must be
performed once every two years and when the Instrument is repaired or when Instrument
parts are replaced.
7HPSHUDWXUH$FFXUDF\
This test checks the performance of the temperature parameter. The temperature accuracy test
is required once every two years.
7RROVUHTXLUHG Patient simulator (with 0.1
o
C or 0.2
o
F)
6WHS Connect the patient simulator to the temperature connector on the Measurement
Server or Measurement Server Extension.
6WHS Configure the patient simulator to 40
o
C or alternatively 100
o
F
6WHS The value should be 40
o
C ± 0.2
o
C or 100
o
F ± 0.4
o
F.
If the displayed value is not within the tolerances given above, refer to "Troubleshooting the
Instrument".
(&*5HVS3HUIRUPDQFH
This test checks the performance of the ECG and respiration parameters. ECG/Resp
performance test is required once every two years.
7RROVUHTXLUHG Patient Simulator.
(&*3HUIRUPDQFH
6WHS Connect the Patient Simulator to the ECG/Resp connector on the Measurement
Server.
6WHS Configure the Patient Simulator as follows:
ECG sinus rhythm.
HR = 100 bpm
6WHS Check the displayed ECG wave and HR value against the simulator configuration.
6WHS The value should be 100bpm +/- 2bpm.
5HVSLUDWLRQ3HUIRUPDQFH
6WHS Change the Patient Simulator configuration to: