User`s guide

M3000A/M3046AM3015A/M3016A Service Guide
4-16 Testing the Instrument
1RLVH&KHFN
6WHS Check that the Monitor is in Service Mode and in the CO
2
window.
6WHS Disconnect the flowmeter and connect the 5% calibration gas and flow regulator in
its place.
6WHS Open the valve to apply the 5% Calibration Gas and wait until the value is stable.
6WHS Check the Noise Index (x6) displayed nect to the CO
2
value on the M3046A dis-
play (this indicates the level of noise on the CO
2
wave). If the value exceeds 3
mmHg the IR Lamp must be replaced.
&2
*DV0HDVXUHPHQW&DOLEUDWLRQ&KHFN
After switching the Measurement Extension on, wait at least 20 minutes before checking the
calibration.
Check the calibration of the CO
2
Gas Measurement as follows:
6WHS Check that the 5% calibration gas and flow regulator are connected.
6WHS Calculate the expected measurement value in mmHg as follows:
0.05 x (ambient pressure) = value mmHg
e.g. 0.05 x 736 = 36.8 mmHg (with an ambient pressure of 736 mmHg)
6WHS Open the valve on the flow regulator to allow 5% CO
2
gas to flow into the Exten-
sion. Allow the value to stabilize.
6WHS Check that the value on the instrument (measurement value on the Main Screen,
x7)) matches the calculated mmHg value ± 2.6 mmHg. If the value is outside the
tolerance, calibrate as described in Step 9 onwards.
6WHS Disconnect the 5% calibration gas and connect the 10% calibration gas.
6WHS Calculate the expected measurement value and tolerance in mmHg as follows:
0.1 x (ambient pressure) = value mmHg
±0.07 x (value mmHg) = tolerance
e.g. 0.1 x 737 mmHg = 73.7 mmHg (with an ambient pressure of 737 mmHg)
±0.07 x 73.7 mmHg = ±5.16 mmHg tolerance
6WHS Open the valve on the flow regulator to allow 10% CO
2
gas to flow into the Exten-
sion. Allow the value to stabilize.
6WHS Check that the value on the instrument (x8) matches the calculated mmHg value
within the calculated tolerance. If so, the Measurement Extension is correctly cali-
brated. If the value is outside the tolerance, calibrate as follows
6WHS If not already connected, connect the 5% calibration gas.
6WHS Select Cal. CO
2
.