User`s guide

M3000A/M3046AM3015A/M3016A Service Guide
4-12 Testing the Instrument
6WHS After these 10 seconds, compare the manometer value with the displayed value.
6WHS Document the value displayed by the M3046A (x3)
6WHS If the difference is greater than 3 mmHg, calibrate the Server (see steps 10 to 12 in
the Accuracy Test procedure).
9DOYH7HV W
6WHS Raise the pressure again to 280 mmHg.
6WHS Select Close valves: Off.
6WHS Wait 5 seconds then document the value displayed. The value should be less than
10 mmHg.
6WHS Document the value displayed by the M3046A (x4).
6LGHVWUHDP&2
3HUIRUPDQFH7HVW
127(
1. Allow 5 seconds between individual service procedures in order to ensure stable
equipment conditions.
2. When certain Monitor procedures are running (for example, AutoZero or purging), service
procedures are not possible and trying to start them will result in a message Service
Operation Failed in the status line on the Monitor. Wait until the Monitor has completed the
current operation , then restart the service procedure.
This test checks the performance of the CO
2
measurement for the sidestream Extension. The
CO
2
performance test is required once per year and when the Instrument is repaired or when
parts are replaced.
7DEOH
Test Expected test results What to record on the Service
record
Accuracy test x1 = value displayed by M3046A
Difference 3mmHg
Leakage test x2 = leakage test value
x2 < 6 mmHg
Linearity test x3 = value displayed by M3046A
Difference 3mmHg
Valve Test x4 = value < 10 mmHg PN:P/x1/x2/x3/x4
or
PN:F/x1/x2/x3/x4
where P = Pass and F = Fail