User`s guide
M3000A/M3046AM3015A/M3016A Service Guide
Testing the Instrument 4-9
6\VWHP6HOI7HVW9DOXHV
0RGXOH 7HVW1XPHULF/LPLWV 7HVW:DYHIRUP
ECG and ECG/Resp
ECG
Resp
100 bpm in ADULT mode
125 bpm in NEO/PEDI mode
15 rpm in ADULT mode
30 rpm in PEDI mode
55 rpm in NEO mode
Simulated squarewave and numeric
Simulated Resp wave and numeric
Pressure
a
Systolic
Diastolic
a. An appropriate transducer must be plugged into the Server or Server Extension in order to get
the test reading.
120 mmHg ADULT
60 mmHg PEDI/NEO
0 mmHg ADULT/PEDI/NEO
Simulated squarewave and numerics
SpO
2
/Pleth
a
SpO
2
Pleth
777
9
100%
Wave on display
Numeric only
Simulated wave
Temperature
a
40 °C (104 °F) Numeric only
Pleth Pulse Rate 60 bpm in ADULT/NEO mode
NBP 120/80 (90) ADULT
100/60 (80) PEDI
80/50 (60) NEO
Simulated squarewave and numerics
Sidestream CO
2
b
b. Appropriate tubing must be plugged into the Server Extension in order to get the test read-
ing. When in Neonatal mode the test signals are switched on, a “CO
2
Equip. Malf.” message
will appear until the test signals are switched off; no action is required.
EtCO
2
40 mmHg
ImCO
2
0 mmHg
AWRR 20 rpm
Simulated squarewave and numerics
Mainstream CO
2
a
EtCO
2
40 mmHg
ImCO
2
0 mmHg
AWRR 20 rpm
Simulated squarewave and numerics