User`s guide

M3000A/M3046AM3015A/M3016A Service Guide
4-6 Testing the Instrument
Serial Numbers
When recording test results, these are always associated with a particular instrument by
means of the serial number. The serial numbers for the Monitor and the Measurement Server
can be seen in the Revisions window (press Setup key then select Revisions). However,
if a Measurement Server Extension (M3015A or M3016A) is in use, the number will not
appear and must be noted down from the back of the Extension. To do this, remove the
Measurement Server and Extension from the Monitor, so that the back of the extension is
visible.
Passwords
The following table show the passwords you need to enter to access the different modes. Go
to Operating Modes and use the TouchStrip to specify the password. Press OK:
Visual Test
Inspect the system for obvious signs of damage. Also check external leads and accessories.
What to record on the service record:
V:P or V: F
where P=pass, i.e. the system has no obvious signs of damage
and F=fail
Power On Test
6WHS Switch on the monitor and connect the M3000A to the M3046A
6WHS Observe whether the monitor boots up successfully without displaying an error
code and if an ECG wave appears on the screen.
What to record on the service record:
'HPR &RQILJ 6HUYLFH
42351 13251 25531