User`s guide

M3000A/M3046A/M3015A/M3016A Service Guide
3-6 Maintaining the Instrument
VHUYLFHSURFHGXUHVDUHQRWSRVVLEOHDQGWU\LQJWRVWDUWWKHPZLOOUHVXOWLQDPHVVDJH
³6HUYLFH2SHUDWLRQ)DLOHG´LQWKHVWDWXVOLQHRQWKH0RQLWRU:DLWXQWLOWKH0RQLWRUKDV
FRPSOHWHGWKHFXUUHQWRSHUDWLRQWKHQUHVWDUWWKHVHUYLFHSURFHGXUH
Preventive maintenance comprises of the following activities:
Perform visual inspection (see Inspecting the Instrument on page 3-4)
Check the operating time for the IR lamp (IR SourceTime), pump and CO
2
scrubbers
(Pump OpTime).
If required, (operating time >15,000 hours), replace the infrared Lamp, the pump and the
CO
2
scrubber, and reset the appropriate operating time counters.
Perform barometric pressure check and calibrate, if necessary, see Barometric Pressure
Check and Calibration on page 4-13.
Perform leakage check, see Leakage Check on page 4-14
Perform Pump Check
Perform flow check and calibrate, if necessary, see Flow Rate Check and Calibration on
page 4-15.
Perform Noise Check
Perform CO
2
calibration check; check the date of the last calibration and calibrate, if neces-
sary, see CO2 Gas Measurement Calibration Check on page 4-16.
If calibration done, perform CO
2
calibration verification using 2nd calibration gas
If M3015A has been opened, perform a system safety test, see Patient Safety Checks on
page 4-24.
7RROV5HT XLUHG
Standard tools, e.g. screwdriver, tweezers
ESD protection equipment
Electronic flowmeter, M1026-60144
Gas calibration equipment
Cal 1 gas 15210-64010 (5% CO
2
)
Cal 2 gas 15210-64020 (10% CO
2
)
Cal gas flow regulator M2267A
Cal tube 13907A
5HTXLUHG5HSODFHPHQW3DUWV