Technical data
91
S:\hp11807a\OPT001\USRGUIDE\BOOK\chapters\descrptn.fb
Chapter 4, Test, Parameter, and Pass/Fail Limit Descriptions
TEST_01. TX and RX Stand-by Current Drain
TEST_01. TX and RX Stand-by Current Drain
This test measures the power-supply current required by the FM radio while it’s in
both the transmit and receive modes. While transmitting, current drain is
measured without modulation, and while receiving, stand-by current drain is
measured after the squelch control is set to maximum.
This test requires that the test set has the optional current-measuring circuit,
Option 003, or that an external power supply is configured to the test set. If an
external power supply is configured, this test reads current from the external
supply.
NOTE: The HP 11807A software does not zero the test set’s current-measuring circuit. Zeroing this
circuit should be done periodically.
The transmitter and receiver stand-by current drain is measured in Amps.
Pass/Fail Limits Used
Pass/Fail Limit 14, RX Stand-by Current Drain
Pass/Fail Limit 27, TX Current Drain