Technical data
54 S:\hp11807a\OPT001\USRGUIDE\BOOK\chapters\softav14.fb
Chapter 2, Using the Software/HP 8920B or HP 8920A FW Above Rev. A.14.00
Customizing Testing
Changing Test Execution Conditions
Test Execution Conditions define where and when test output occurs. You
may decide to:
• Display output on CRT only, or display on CRT and print hardcopy (Output
Results To).
NOTE: If printing test results is desired, after selecting Printer, additional steps are necessary to
connect and configure the printer. See "Printing," in chapter 5, on page 212.
• Display (or print) only measurements that fail, or display (or print) all measurements
that pass or fail (Output Results For).
• Enter a title for an output heading for the displayed or printed results (Output
Heading).
• Stop testing when a measurement fails or continue through all of the tests without
stopping (If Unit-Under-Test-Fails).
• Pause between each measurement, or run through entire test (Test Procedure
Run Mode).
• Start the program automatically when the Test System is powered on. (Autostart
Test Procedure on Power-up)
Test Execution Conditions
is accessed from the SETUP TEST SET: list.
To change a default setting, position the cursor to the desired field. Pressing the
knob (”selecting”) will toggle the underlined selection.
Test Execution Conditions settings are not retained after a power-down/
power-up cycle, and will return to their default settings.