Technical data

34 S:\hp11807a\OPT001\USRGUIDE\BOOK\chapters\softav14.fb
Chapter 2, Using the Software/HP 8920B or HP 8920A FW Above Rev. A.14.00
Customizing Testing
Changing the Order of Tests
You may define the order of tests to include all, some, or just one of the tests
available. When the first test is finished, the next will run. The test sequence will
remain in the test set’s battery backed-up memory until another test sequence is
loaded or set up. For information on saving a customized test sequence, see
"Saving a Test Procedure" on page 50.
Defining the order of tests is accomplished by inserting or deleting tests from the
list of tests that come with the software package.
See chapter 4, "Test, Parameter,
and Pass/Fail Limit Descriptions" on page 89
, for descriptions of tests included in
this package.
The
All Chans? field allows the user to decide to run the test on all channels
entered in the
Channel Information table, or just the channels which are
selected as
Prime in the Channel Information screen. This feature allows the
user the flexibility to use channels that are selected as
Prime in all the tests in the
sequence, and those indicated as non-prime in a subset of tests (those tests with a
Yes response in All Chans). For more information, see "Specifying Channel
Information" on page 38
.
The following describes how to create a new test sequence and enter a response to
the
All Chans? field.