Technical data
234
S:\hp11807a\OPT001\USRGUIDE\BOOK\chapters\referenc.fb
Chapter 5, Reference (Alphabetical)
Test Execution Conditions
Test Execution Conditions
In some situations, you may wish to change the way the software works when a
test result is obtained.
Test Execution Conditions allow you to do this.
Test Execution Conditions are accessed from the SET UP TEST SET list
on the TESTS (Main Menu) screen. Press TESTS, then select
Execution Cond
to display them. (They are displayed directly on the main TESTS screen in
firmware below revision A.14.00).
Test Execution Conditions are not retained after a power-down/power-up
cycle.
The following
Test Execution Conditions can be set as needed:
Output Results To: (Output Destination)
You can select either: Crt or Printer Default:Crt
You can specify where test results are to be placed. If you select
Crt, results will
be displayed on the test set’s CRT. If you select
Printer, test results will be sent
to the CRT and to a printer. You must connect and configure a printer if you select
Printer. See "Printing" on page 212.
Output Results For: (Output Results)
You can select either: All or Failures Default:All
You can specify if you want only the failed results to be displayed or printed.
This will be useful if you generally do not print test results, and want to ensure
that failed results are displayed or printed.
Output Heading
You can use this field to enter a heading that will be printed or displayed.
Select the
Output Heading field. Use the knob to choose and enter the
characters that you want to appear in the heading.
Choose
Done when you are finished.