Technical data
 205
S:\hp11807a\OPT001\USRGUIDE\BOOK\chapters\referenc.fb
Chapter 5, Reference (Alphabetical)
Memory Cards
Memory Card Storage Space 
Procedures use 12-16 records each. A Library uses 20-35 records. A single library 
must be included on the card. A record is 256 bytes. Approximately 11 kilobytes 
of overhead is required on each card. 
Use the following formula to estimate the storage space needed: 
Storage Space(in kilobytes) = (Number of Procedures × 4.1) + 20
For example, if you want to save ten different procedures, you will need 61 
kilobytes of memory. The 64 kilobyte or 128 kilobyte card is sufficient. 
The storage space you need for data collection depends on the number of test 
results that are saved. You will need approximately 4 kilobytes per page of test 
results that you save. A page of test results is about 57 lines of CRT or printer 
output. 
The storage space of smaller SRAM cards can be quickly used. If you are 
collecting large quantities of data, data collection using a PC or printer may be 
preferable. 
Initializing a Memory Card 
Initializing HP 11807A cards (Smart Cards) using the TESTS Save/Delete 
Procedure
 screen automatically defaults to LIF format. Initializing HP 11807E 
cards (PCMCIA) using the TESTS 
Save/Delete Procedure screen 
automatically defaults to DOS format. However, initializing cards from the 
Save/Delete Procedure screen is only available on HP 8920B or HP 
8920A,D test sets with firmware above revision A.14.00. If these settings do not 
match your needs, or you have firmware below revision A.14.00, there is another 
method described below in which you may select the format. 
• Press TESTS. 
• Select Save/Delete Procedure from the CUSTOMIZE TEST PROCEDURE 
list. 
• Insert the SRAM card in the slot on the front panel. (Make sure the switch on the card 
is not in the write-protected position). 
•Press k3 (Init Card). 
• Press Yes if you want to continue. 