Technical data
175
S:\hp11807a\OPT001\USRGUIDE\BOOK\chapters\descrptn.fb
Chapter 4, Test, Parameter, and Pass/Fail Limit Descriptions
Pass/Fail Limit 31, TX Microphone Sensitivity
Pass/Fail Limit 31, TX Microphone Sensitivity
This sets the pass/fail limits for testing a transmitter’s microphone sensitivity.
Microphone sensitivity is the nominal voltage measured at the input of the
microphone which will cause full-rated system deviation of the transmitter. The
value for full-rated system deviation is set by
Parameter 2, RT Full Rated System
Deviation
.
Pass/Fail limits are determined from an applicable radio specification or industry
standard, such as:
• Radio specification: Transmitter Audio Sensitivity
• Industry standard: EIA/TIA-603, Section 2.2.10, Acoustic Microphone Sensitivity
For Example:
If your radio specification defined the microphone sensitivity to be 100 mVrms
±10%, enter 90 in the lower-limit column and 110 in the upper-limit column.
Limits are entered in mVrms.
It’s Used in the Following Tests:
TEST_07. TX Microphone Sensitivity
TEST_21. TX Quick Test
This pass/fail limit is used as a reference value whenever transmitter deviation is
set during testing.