Chapter 4, Test, Parameter, and Pass/Fail Limit Descriptions
Pass/Fail Limit 27, TX Current Drain
Pass/Fail Limit 27, TX Current Drain
This sets the pass/fail limits for testing a transmitter’s current drain.
Pass/Fail limits are determined from an applicable radio specification, such as:
Radio pass/fail limits; Maximum Battery Drain, Transmit at Rated Power
For Example:
If your radio specification defines the maximum battery drain while transmitting
as 5 Amps and you want to test the radio’s upper limits, enter
0 in the lower-limit
column and
5 in the upper-limit column. Limits are entered in Amps.
It’s Used in the Following Tests:
ALL TESTS. Whenever a system initialization is done.
TEST_01. TX and RX Stand-by Current Drain