Chapter 4, Test, Parameter, and Pass/Fail Limit Descriptions
Pass/Fail Limit 17, RX Usable Sensitivity
Pass/Fail Limit 17, RX Usable Sensitivity
This sets the pass/fail limits for testing usable sensitivity (in TEST_14. RX Usable
Sensitivity
and TEST_20. RX Quick Test) and for testing the variation of sensitivity
with signal frequency changes (in
TEST_19. RX Variation of Sens with Freq).
Pass/Fail limits are determined from an applicable radio specification or industry
standard, such as:
• Radio specification: Sensitivity
• Industry standard: EIA/TIA-603, Section 2.1.4 Reference (Usable) Sensitivity,
• and EIA/TIA-204-D, Section 8.0, Variation of Sensitivity with Signal Frequency
For Example:
If your radio specification defined sensitivity to be 1
µV and you want to test the
radio’s upper limits, you would enter
0 in the lower-limit column and 1 in the
upper-limit column. Limits are entered in
µV.
It’s Used in the Following Tests:
TEST_14. RX Usable Sensitivity
TEST_19. RX Variation of Sens with Freq
TEST_20. RX Quick Test