Technical data

147
S:\hp11807a\OPT001\USRGUIDE\BOOK\chapters\descrptn.fb
Chapter 4, Test, Parameter, and Pass/Fail Limit Descriptions
Pass/Fail Limit 3, RX Audio Freq Resp Delta from 6 dB/Oct
Pass/Fail Limit 3, RX Audio Freq Resp Delta from 6 dB/Oct
This defines pass/fail limits for testing a receiver’s audio-frequency response,
with squelch set to minimum and volume set to 50% of rated audio-output power.
Pass/Fail limits are determined from an applicable radio specification or industry
standard, such as:
Radio specification: Audio Response
Industry standard: EIA/TIA-603, Section 2.1.10, Audio Frequency Response.
For Example:
If your radio specification defines receiver audio response to be within +1 and
3 dB of a 6 dB/octave de-emphasis curve and you want to test both limits on the
radio, enter
3 in the lower-limit column and 1 in the upper-limit column. Limits
are entered in dB.
It’s Used in the Following Tests:
TEST_13. RX Audio Frequency Response
TEST_20. RX Quick Test