Technical data
130 S:\hp11807a\OPT001\USRGUIDE\BOOK\chapters\descrptn.fb
Chapter 4, Test, Parameter, and Pass/Fail Limit Descriptions
Parameter 18, RX Usable Sens Set and Measure
Parameter 18, RX Usable Sens Set and Measure
This test parameter allows you to choose how the test for usable sensitivity will be
performed.
For Example:
Setting this test parameter to “
1” causes the software to perform a pass/fail
measurement of the receiver’s usable sensitivity. In this case, the test set RF
generator’s output level is set to the upper limit of
Pass/Fail Limit 13, RX Squelch
Blocking
for “RX usable sensitivity” and the measured SINAD level at the radio’s
output is checked against test
Parameter 18, RX Usable Sens Set and Measure for
“RX usable sensitivity SINAD level”. If the SINAD level is within the specified
limits the test passes and the specified usable sensitivity from
Pass/Fail Limit 13,
RX Squelch Blocking
is displayed in the “Measured value” column.
Setting this test parameter to “
0” causes the software to perform a faster, iterative
measurement of usable sensitivity. In this case, the test set’s RF generator’s output
level is set to the “RX Standard RF Input Level” (determined by the value in test
Parameter 15, RX Squelch Preset Only
). The RF generator’s output level is varied
until the SINAD level measured at the radio’s output. The SINAD level measured
is displayed in the “
Measured value” column as the usable sensitivity SINAD
level.
It’s Used in the Following Tests:
TEST_14. RX Usable Sensitivity
TEST_20. RX Quick Test