Technical data

Service and Maintenance 8
Recommended Test Equipment
E364xA User’s and Service Guide 189
Recommended Test Equipment
Table 8- 4 describes the test equipment recommended for the
performance verification and adjustment procedures. If the
exact instrument is not available, use the accuracy
requirements shown to select substitute calibration
standards. If you use equipment other than the
recommended equipment shown in Table 8- 4, you must
recalculate the measurement uncertainties for the actual
equipment used.
Table 8 - 4 Recommended test equipment
Equipment Requirements Recommended model Test function
GPIB controller Full GPIB or RS-232 capabilities Agilent 82357B interface
card or equivalent
Programming and readback
accuracy
Oscilloscope 100 MHz with 20 MHz bandwidth Agilent DSO5034A or
equivalent
Display transient response
and ripple and noise
waveform
Rms voltmeter 20 Hz to 20 MHz Agilent 54602B oscilloscope
or R&S URE3 rms voltmeter
Measure rms ripple and noise
Cable (BNC-to-BNC) 50 Ω, 9 inch (23 cm) Agilent 10502A or 10503A if
the 10502A is not available
Measure rms ripple and noise
(CV PARD, CC PARD)
BNC (female) bulkhead
receptacle
Isolated ground
Nominal impedance: 50 Ω
Pomona Model 5148 Measure rms ripple and noise
(CV PARD, CC PARD)
Split ferrites For use with round cable Steward Co. 28A2029-0A0 Noise coupling reduction
Digital voltmeter Resolution: 0.1 mV
Accuracy: 0.01%
Agilent 34401A Measure DC voltages
Electronic load Voltage range: 240 Vdc
Current range: 10 Adc
Open and short switches
Transient on/off
Agilent 6060B or equivalent Measure load and line
regulations and transient
response time
Resistive loads (R
L
) (2.7 Ω, 150 W/13.5 Ω, 150 W)
[1]
(43.8 Ω, 300 W/120 Ω, 300 W)
[2]
(1.6 Ω, 300 W/8.0 Ω, 300 W)
[3]
(25 Ω, 300 W/75 Ω, 300 W)
[4]
Measure ripple and noise