Technical data

Remote Interface Reference 3
Output Settings and Operation Commands
E364xA User’s and Service Guide 91
VOLTage:PROTection:CLEar
Cause the overvoltage protection circuit to be cleared. After
this command, the output voltage is restored to the state it
was in before the protection feature occurred and the OVP
trip level remains unchanged to the value presently
programmed. Before sending this command, lower the output
voltage below the OVP trip point, or raise the OVP trip level
above the output setting. Note that the overvoltage condition
caused by an external source must be removed first before
proceeding with this command.
VOLTage:RANGe
{P8V
[1]
|P20V
[1]
|P35V
[2]
|P60V
[2]
|LOW|HIGH}
Select an output range to be programmed by the identifier.
For example, P20V or HIGH is the identifier for the 20 V/1.5
A range and P8V or LOW is for the 8 V/3 A range (for the
E3646A model). At *RST, the low voltage range is selected.
VOLTage:RANGe?
Query the currently selected range. The returned parameter
is P8V or P35V for the low voltage range, or P20V or P60V
for the high voltage range.
[1] Applicable for E3646A and E3648A only.
[2] Applicable for E3647A and E3649A only.