Technical data
3 Remote Interface Reference
Output Settings and Operation Commands
90 E364xA User’s and Service Guide
VOLTage:TRIGgered? [MINimum|MAXimum]
Query the triggered voltage level presently programmed. If
no triggered level is programmed, the VOLT level is returned.
VOLTage:PROTection
{<voltage>|MINimum|MAXimum}
Set the voltage level at which the overvoltage protection
(OVP) circuit will trip. If the peak output voltage exceeds the
OVP level, then the power supply output is shorted by an
internal SCR. An overvoltage condition can be cleared with
the VOLT:PROT:CLE command after the condition that
caused the OVP trip is removed.
VOLTage:PROTection? [MINimum|MAXimum]
Query the overvoltage protection trip level presently
programmed.
VOLTage:PROTection:STATe {0|1|OFF|ON}
Enable or disable the overvoltage protection function. At
*RST, this value is set to ON.
VOLTage:PROTection:STATe?
Query the state of the overvoltage protection function. The
returned parameter is 0 (OFF) or 1 (ON).
VOLTage:PROTection:TRIPped?
Return a 1 if the overvoltage protection circuit is tripped
and not cleared or a 0 if not tripped.