Specifications
This series of 2000 watt DC power
supplies has the exceptional, proven
reliability that test system engineers
look for. It also has the unusual
combination of high efficiency
and low noise operation.
Programming of the DC output and
the extensive protection features
can be done either from the front
panel or using industry standard
SCPIcommands,viatheGPIB.
Using the serial link, up to 16 power
supplies can be connected through
oneGPIBaddress.Testsystem
integration can be further simplified
beusingtheVXIPlug&Play drivers.
The output voltage and current
can also be controlled with analog
signals. This is helpful for certain
types of noisy environments,
and also immediate reactions
to process changes.
Lab-bench use is enhanced by the
fan-speed control, which minimizes
acoustic noise. The extremely low
ripple and noise helps the built-in
measurement system make extreme-
ly accurate current and voltage
measurements.
High Performance DC Power Supplies speed and accuracy for test optimization
Specifications
(at 0° to 55°C unless
otherwise specified)
More detailed specifications at
www.agilent.com/find/6670
79
Fast, low-noise outputs
Analog control of output voltage and current
Fan-speed control to minimize acoustic noise
Built-in measurements and advanced programmable features
Protection features to ensure DUT safety
6671A - 6675A
Single-Output
2000 W GPIB
6671A 6672A 6673A 6674A 6675A
Number of outputs
1 1 1 1 1
GPIB
Yes Yes Yes Yes Yes
Output ratings
Output voltage 0 to 8 V 0 to 20 V 0 to 35 V 0 to 60 V 0 to 120 V
Output current 0 to 220 A 0 to 100 A 0 to 60 A 0 to 35 A 0 to 18 A
Programming accuracy
at 25°C ±5°C
Voltage 0.04% + 8 mV 20 mV 35 mV 60 mV 120 mV
Current 0.1% + 125 mA 60 mA 40 mA 25 mA 12 mA
Ripple and noise
from 20 Hz to 20 MHz
Voltage rms 650 V 750 V 800 V 1.25 mV 1.9 mV
Voltage peak to peak 7 mV 9 mV 9 mV 11 mV 16 mV
Current rms 200 mA 100 mA 40 mA 25 mA 12 mA
Readback accuracy at 25°C ±5°C
(percent of reading plus fixed)
Voltage 0.05% + 12 mV 30 mV 50 mV 90 mV 180 mV
±Current 0.1% + 150 mA 100 mA 60 mA 35 mA 18 mA
Load and line regulation
Voltage 0.002%+ 300 V 650 V 1.2 mV 2 mV 4 mV
Current 0.005%+ 10 mA 7 mA 4 mA 2 mA 1 mA
Transient response time Less than 900 s for the output voltage to recover 100 mV following a
change in load from 100% to 50% or 50% to 100% of the output current
rating of the supply
Supplemental Characteristics
(Non-warranted characteristics determined by design and
useful in applying the product)
Average resolution
Voltage 2 mV 5 mV 10 mV 15 mV 30 mV
Current 55 mA 25 mA 15 mA 8.75 mA 4.5 mA
OVP 15 mV 35 mV 65 mV 100 mV 215 mV
Output Voltage programming
response time*
(excluding command 30 ms 60 ms 130 ms 130 ms 195 ms
processing time)
* Full load programming rise/fall time (10% to 90% or 90% to 10%) with full resistive load equal to rated output
voltage/rated output current.