Specifications

65
Accurately characterize your
DUT’s power profile with advanced
measurements
The APS provides simultaneous
voltage and current measurement
capabilities that deliver high
accuracy and resolution. Make mea-
surements using two main modes:
averaged or digitized. In average
mode, the APS delivers high-accuracy
DMM-quality voltage and current
measurements. The digitizing
capability allows you to capture
dynamic current or voltage profiles
at much higher resolution than an
oscilloscope.
Capture inrush current: APS
provides an 18-bit current digitizer
with a sample rate up to 200 kS/s,
level triggers, and a current
measurement range that is 2.25x
higher than the max output cur-
rent range of the power supply.
Single Output: 1 & 2 KW GPIB, LAN (LXI Core), USB
(Continued)
Specifications
(at to 55°C unless
otherwise specified)
More detailed specifications at
www.agilent.com/find/N6900
Advanced Power System DC power supplies overcome the toughest power test challenges
N6950A N6951A N6952A N6953A N6954A
Supplemental Characteristics (Non-warranted characteristics determined by design
that are useful in applying the product)
Volt up-programming time
4
10% to 90% 3 ms 3 ms 3 ms 3 ms 3 ms
Settling time
6
10 ms 10 ms 10 ms 10 ms 10 ms
Volt down-programming time
5
90% to 10% 3 ms 3 ms 3 ms 3 ms 3 ms
Settling time
6
10 ms 10 ms 10 ms 10 ms 10 ms
Current up-programming time
7
10% to 90% 2.5 ms 2.5 ms 2.5 ms 2.5 ms 2.5 ms
Output-on delay time
Voltage priority 12 ms 12 ms 12 ms 12 ms 12 ms
Current priority 14 ms 14 ms 14 ms 14 ms 14 ms
Line regulation
8
Voltage <10 µV <10 µV <10 µV <10 µV <10 µV
Current <10 µA <10 µA <10 µA <10 µA <10 µA
Notes:
4
With full resistive load and a voltage transition from 0.1% to 100% of rated output
5
With no load and a voltage transition from 100% to 0.1% of rated output
6
From start of voltage change to within 0.1% of final full scale value
7
With full resistive load and a current transition from 0.1% to 100% of rated output
8