Specifications
This series of linear-regulated
40-50 W DC power supplies is
designed to maximize the
throughput of DUTs through the
manufacturing test process with
fast programming and measurement,
and also active downprogramming.
It offers many advanced program-
mable features including stored
states and status reporting. Pro-
gramming is done using industry
standard SCPI commands via the
GPIBorRS-232.Testsystem
integration is further simplified by
usingtheVXIPlug&Play drivers.
The optional relays simplify system
design and troubleshooting.
The half-rack size of the 6610C
series makes it a convenient
DC power supply for the R&D lab
bench. The built-in microamp
measurement system helps the
engineer to easily and accurately
monitor the output voltage and
current without a complicated
test setup.
Application notes
10 Practical Tips You Need to
Know About Your Power Products
5965-8239E
10 Hints for Using Your Power Supply
to Decrease Test Time
5968-6359E
Understanding Linear Power Supply
Operation (AN1554)
5989-2291EN
High Performance DC Power Supplies speed and accuracy for test optimization
More detailed specifications at
www.agilent.com/find/6610
45
Small, compact size for bench and system use
Fast, low-noise outputs
Dual-range, precision low current measurement
Built-in measurements and advanced programmable features
Protection features to ensure DUT safety
6611C - 6614C
Single-Output
40-50 W GPIB
Specifications
(at 0° to 55°C unless
otherwise specified)
6611C 6612C 6613C 6614C 6611C-
J05
Special order
option
Number of outputs 1 1 1 1 1
GPIB Yes Yes Yes Yes Yes
Output ratings
Voltage 0 to 8 V 0 to 20 V 0 to 50 V 0 to 100 V 0 to 10 V
Current 0 to 5 A 0 to 2 A 0 to 1 A 0 to 0.5 A 0 to 5 A
Programming accuracy (at 25°C ±5°C)
Voltage 0.05% + 5 mV 10 mV 20 mV 50 mV 5 mV
+Current 0.05% + 2 mA 1 mA 0.75 mA 0.5 mA 2 mA
Ripple and noise 20 Hz to 20 MHz, with outputs
ungrounded or with either terminal grounded
Voltage rms 0.5 mV 0.5 mV 0.5 mV 0.6 mV 0.5 mV
peak -to-peak 3 mV 3 mV 4 mV 5 mV 3 mV
Normal mode rms 2 mA 1 mA 1 mA 1 mA 2 mA
DC measurement accuracy via GPIB or front-panel
meters with respect to actual output at 25°C ±5°C
Voltage 0.03% + 2 mV 3 mV 6 mV 12 mV 2 mV
Low current range -20 mA to + 20 mA 0.1% + 2.5 A 2.5 A 2.5 A 2.5 A 2.5 A
High current range +20 mA to + rated I 0.2% + 0.5 mA 0.25 mA 0.2 mA 0.1 mA 0.5 mA
-20 mA to - rated I
0.2% + 1.1 mA 0.85 mA 0.8 mA 0.7 mA 1.1 mA
Load regulation
Voltage 2 mV 2 mV 4 mV 5 mV 2 mV
Current 1 mA 0.5 mA 0.5 mA 0.5 mA 1 mA
Line regulation
Voltage 0.5 mV 0.5 mV 1 mV 1 mV 0.5 mV
Current 0.5 mA 0.5 mA 0.25 mA 0.25 mA 0.5 mA
Transient response time Less than 100 s for the output to recover to its previous level (within 0.1% of the voltage
rating of the supply or 20 mV, whichever is greater) following any step change in load current of up to 50% of the
output current rating of the supply
Supplemental Characteristics
(Non-warranted characteristics determined by design and
useful in applying the product)
Average programming resolution
Voltage 2 mV 5 mV 12.5 mV 25 mV 3 mV
Current 1.25 mA 0.5 mA 0.25 mA 0.125 mA 1.25 mA
Sink current -3 A -1.2 A -0.6 A -0.3 A -3 A