Performance specification (continued)
Voltage readback
12 months (over USB with respect to the actual output at 25 °C ± 3 °C), ±(% of output + offset)
Accuracy
1
Range ±2 V 0.075% + 1.5 mV 0.075% + 1.5 mV
Range ±20 V 0.05% + 10 mV 0.05% + 10 mV
Resolution Range ±2 V 0.1 mV 0.1 mV
Range ±20 V 1 mV 1 mV
Current readback 12 months (over USB with respect to the actual output at 25 °C ± 3 °C), ±(% of output + offset)
Accuracy
1
Range ±1 µA 0.085% + 0.85 nA 0.085% + 0.85 nA
Range ±10 µA 0.085% + 8.5 nA 0.085% + 8.5 nA
Range ±100 µA 0.075% + 75 nA 0.075% + 75 nA
Range ±1 mA 0.075% + 750 nA 0.075% + 750 nA
Range ±10 mA 0.075% + 7.5 µA 0.075% + 7.5 µA
Range ±120 mA 0.1% + 100 µA 0.1% + 100 µA
Resolution Range ±1 µA 100 pA 100 pA
Range ±10 µA 1 nA 1 nA
Range ±100 µA 10 nA 10 nA
Range ±1 mA 100 nA 100 nA
Range ±10 mA 1 µA 1 µA
Range ±120 mA 20 µA 20 µA
Rise/fall time (ms)
2
For resistive measurement
2
±1 µA 170.0 15.0
±10 µA 18.0 5.0
±100 µA 6.0 1.0
±1 mA 1.0 1.0
±10 mA 1.0 1.0
±120 mA 1.0 1.0
Specifications
(at 0° to 55°C unless
otherwise specified)
U2722A U2723A
143
Pre-define test configurations and
execute commands automatically
TheU2723AUSBmodularsource
measure unit provides an embedded
test script to help you pre-define
test configurations or duplicate
tests easily without spending
too much time on programming.
EachchannelintheU2723AUSB
modular SMU is allocated two
memory lists, each capable of stor-
ing up to 200 commands and results
individually. Stored commands in
active memory will be executed
accordingly while the measurement
results obtained are automatically
stored in the result buffer.
Supplemental characteristics
Remote interface:
•Hi-SpeedUSB2.0*
•USBTMC-USB488
1
Power consumption:
• +12 VDC, 3 A maximum
• Isolated ELV supply source
Operating environment:
• Operating temperature from
0 °C to +50 °C
• Relative humidity at 20% to 85%
RH (non-condensing)
• Altitude up to 2000 meters
• Pollution Degree 2
• For indoor use only
Storage compliance: –20 °C to 70 °C
Safety compliance
Certified with:
• IEC 61010-1:2001/EN 61010-1:2001
(2nd Edition)
• USA: ANSI/UL 61010-1:2004
• Canada: CSA C22.2 No.61010-1:2004
EMC compliance:
• IEC 61326-1:2005/EN61326-1:2006
• Canada: ICES-001:2004
• Australia/New Zealand:
AS/NZS CISPR 11:2004
1
Compatible with Microsoft® Windows®
operating systems only.
*
If remote connections are necessary, a E5813A
USB/LAN hub can be used. Please go to the
product’s user guide for more information.
Source Measure Units
U2700A Series USB Modular Source Measure Units (Continued)
Application Specific DC Power Supplies tailored solutions for specific needs
1
Accuracy measurments are based on NPLC 10.
2
Drive 50% of 1 V or 10 V output with a resistive load. Rise time is from 10% to 90% of program voltage
change at maximum current. Fall time is from 90% to 10% of program voltage change at maximum current.