Specifications
134
Fast I-V Curve Changes
The E4360 offers fast curve changes
to enable better simulation of solar
arrays under various environmental
conditions, like eclipse and spin.
The resolution of the I-V curve can
be set to optimize the I-V curve for
resolution or fast curve change. In
simulation mode and table mode,
you can select high resolution which
uses a 4,096 point table to generate a
smoother I-V curve within 250 msec.
For fast I-V curve generation, you
can select the 256 point table that
quickly generates an I-V curve within
30 msec. All the E4360 SAS in the
system can be synchronized to change
their I-V curves at the same time
using the hardware trigger, such
that I-V curves can be changed on
up to 100 outputs within 30 msec
or 250 msec based on resolution
setting.
I-V Curve List
The E4360 offers a LIST mode that
lets you pre-program a LIST of up
to 512 I-V curves. Program up to
512 sets of points, where each set
of points include curve parameters:
V
oc,
V
mp,
I
sc,
and I
mp
. A dwell time
of 30 ms to 65 seconds with 1 ms
resolution can be specified and the
E4360 SAS will stay (i.e. dwell) at
the set point for the programmed
dwell time value. Alternately, the
LIST can be paced (advanced to the
next set point) by a bus trigger or
it can be paced by a trigger signal
which enables synchronization of
the LIST with an external event.
Utilizing I-V curve lists speed up
test execution by removing the
computer I/O from the process
and simplifies I-V curve change
more easily simulating the solar
array under various conditions.
Modular Solar Array Simulators:
1200 W
(Continued)
More detailed specifications at
www.agilent.com/find/E4360
Application Specific tailored solutions for specific needs
Interface Capabilities
GPIB,SCPI-1993,IEEE488.2
compliant interface
LXICompliance,Core(onlyapplies
tounitswithLXIlabelonfrontpanel)
USB2.0,RequiresAgilentIOLibrary
version M.01.01 and up, or 14.0 and up
10/100 LAN, Requires Agilent IO Library
version L.01.01 and up, or 14.0 and up
Built-inWebserver,RequiresInternet
Explorer 5+ or Netscape 6.2+
Environmental Conditions
Operating environment: Indoor use,
installation category II (for AC input),
pollution degree 2
Temperature range: 0°C to 55°C
(current is derated 1% per °C
above 40°C ambient temperature)
Relative humidity: Up to 95%
Altitude: Up to 2000 meters
Storage temperature: -30°C to 70°C
LED statement: Any LEDs used
in this product are Class 1 LEDs
as per IEC 825-1
Calibration Interval
1 year
Regulatory Compliance
EMC:
Complies with the European EMC
directive 89/336/EEC for Class A test
and measurement products.
Complies with the Australian standard
and carries the C-Tick mark.
This ISM device complies with Canadian
ICES-001. Cet appareil ISM est conforme
àlanormeNMB-001duCanada.
Electronic discharges greater than 1 kV
near the I/O connectors may cause the unit
to reset and require operator intervention.
Safety
Complies with the European Low Voltage
Directive 73/23/EEC and carries the
CE-marking.
This product also complies with the US
and Canadian safety standards for test
and measurement products.
Acoustic Noise Declaration
This statement is provided to comply with
the requirements of the German Sound
Emission Directive, from 18 January 1991.
SoundPressureLp<70dB(A),*At
Operator Position, *Normal Operation,
*According to EN 27779 (Type Test).
SchalldruckpegelLp<70dB(A),*Am
Arbeitsplatz,*NormalerBetrieb,*Nach
EN 27779 (Typprüfung).
Output Terminal Isolation
±240 VDC (maximum from chassis ground)
Dimensions
Height 88.9 mm (3.5 in.)
Width 432.5 mm (17.03 in.)
Depth 633.9 mm (24.96 in.)
Weight
Mainframe with 2 modules:
38.4 lbs (17.6 kg)
Single output module:
7.2 lbs (3.3 kg)
Supplemental Characteristics