Specifications
66000 Modular Power System
The Agilent 66000 modular power
system simplifies test-system
assembly, cabling, programming,
debugging and operation. It is
ideal for ATE and production test
environments, where it can supply
bias power and stimulus to subas-
semblies and final products.
The modular power system
saves rack space, the 7-inch-high
(4-EIA units) mainframe can
accommodate up to eight DC
power modules.
Key features
•GPIB-programmablevoltage
and current
•Programmableover-voltage
and over-current protection
•Self-testinitiatedatpower-up
orfromGPIBcommand
•Electroniccalibrationover
GPIBorfromkeyboard
•Over-temperatureprotection
•Discretefaultindicator/remote
inhibit (DFI/RI)
•Fivenonvolatilestore-recall
states per output
•User-definablepower-onstate
Multiple Mainframes at
One GPIB Address
The Agilent serial link feature will
allow you to control up to 16 outputs
atoneGPIBaddressbyconnecting
an auxiliary mainframe. The serial
link cable comes standard with the
High Performance DC Power Supplies speed and accuracy for test optimization
Specifications
(at 0° to 55°C unless
otherwise specified)
66101A 66102A 66103A 66104A 66105A 66106A
More detailed specifications at
www.agilent.com/find/66000
114
Modular system permits up to 8 outputs of 150 W per output in 4U of rack space
Reconfigure fast with easily swappable modules
Fast, low-noise outputs
LIST mode and advance triggering system
Optional isolation and polarity reversal relays
Built-in measurements and advanced programmable features
Protection features to ensure DUT safety
66000A (mainframe)
66001A (keyboard)
Output ratings at 40°C
Output voltage 0 to 8 V 0 to 20 V 0 to 35 V 0 to 60 V 0 to 120 V 0 to 200 V
Output current 0 to 16 A 0 to 7.5 A 0 to 4.5 A 0 to 2.5 A 0 to 1.25 A 0 to 0.75 A
Maximum power 128 W 150 W 150 W 150 W 150 W 150 W
Programming accuracy
at 25°C ±5°C
Voltage 0.03% + 3 mV 8 mV 13 mV 27 mV 54 mV 90 mV
Current 0.03% + 6 mA 3 mA 2 mA 1.2 mA 0.6 mA 0.4 mA
Readback accuracy
(via GPIB or keyboard
display at 25°C ±5°C)
Voltage 0.02%+ 2 mV 5 mV 8 mV 16 mV 32 mV 54 mV
Current 0.02%+ 6 mA 3 mA 2 mA 1 mA 0.6 mA 0.3 mA
Ripple and noise (20 Hz to 20 MHz)
Constant voltage rms 2 mV 3 mV 5 mV 9 mV 18 mV 30 mV
peak-peak 5 mV 7 mV 10 mV 15 mV 25 mV 50 mV
Constant current rms 8 mA 4 mA 2 mA 1 mA 1 mA 1 mA
Line regulation
Voltage 0.5 mV 0.5 mV 1 mV 2 mV 3 mV 5 mV
Current 0.75 mA 0.5 mA 0.3 mA 0.1 mA 50 A 30 A
Load regulation
Voltage 1 mV 1 mV 1 mV 2 mV 4 mV 7 mV
Current 0.5 mA 0.2 mA 0.2 mA 0.1 mA 50 A 30 A
Transient response time Less than 1 ms for the output voltage to recover within 100 mV of its
previous level following any step change in load current up to 10 percent
of rated current
Supplemental Characteristics (Non-warranted characteristics determined by design
that are useful in applying the product)
Average resolution
Voltage 2.4 mV 5.9 mV 10.4 mV 18.0 mV 36.0 mV 60.0 mV
Current 4.6 mA 2.3 mA 1.4 mA 0.75 mA 0.39 mA 0.23 mA
Output voltage programming (OVP) 50 mV 120 mV 200 mV 375 mV 750 mV 1.25 mV
OVP accuracy 250 mV 500 mV 800 mV 1 V 1.5 V 2.5 V
Modular Power System
1200 W per mainframe GPIB