Specifications

98 Calibration and Preventive Maintenance Chapter 6
System Adjustments
Agilent E6171A/B Measurement Control Module adjustment must be done
in the Agilent TS-5400 Test System where the E6171A/B module resides.
Agilent E6171A/B specific measurement constants are stored in the PC
controller as part of the TestExec SL software. A “soft” or system-wide
adjustment can be done at any time. The TestExec SL software notifies you
if 30 days have expired since the last system-wide adjustment. A complete
module calibration should be performed annually or when anything in the
system changes (for example, moving the system to a new location,
changing any VXI module (Agilent E6171A/B or E1411B DMM, etc.). The
Agilent E1411B in the test system is the reference and must be calibrated
first.
Procedure Do the following:
1. Select and run the ‘autoadj.tpa’ testplan to perform the
system-wide adjustment.
2. Locate themanual’ entry in the AutoAdj action in the “Parameters”
section of the ‘
autoadj.tpa’ testplan form.
3. Set manual to 0 (default) to only perform the system-wide
adjustment; set
manual to 1 to perform both the system-wide
adjustment AND the Agilent E6171A/B manual adjustments.
4. When the ‘autoadj.tpa’ testplan executes the Test AutoAdj
Query
test, the system prompts for a ‘YES’ or ‘NO’ response. Enter
YES to execute the Test Auto Adjust action.
Note Remove any test fixture before running ‘autoadj.tpa’ or the autoadj
action (action is part ofautoadj.tpa’).