Specifications

System Instrument Configuration 39Chapter 2
VXI Technology
E1563A 2-Channel
800 ka/s Digitizer
(ADC) Module
The Agilent E6183A contains an VXI Technology E1563A Digitizer and
cables. The E1563A Digitizer has two channels that allows simultaneous
measurements of two waveforms (for example, the voltage and current
waveform of the load signal). In the factory default configuration, Channel
1 provides direct access to the attenuator output. Connecting the digitizer to
this input allows the user to digitize a high voltage like an inductive flyback,
while maintaining the resolution to also measure the driver’s saturation
voltage all in a single measurement cycle for high-throughput testing.
Channel 2 input is connected to the interconnection column (ABus).
Connecting the digitizer to this input allows it to be routed to any input of
the UUT.
Switch/Jumper Settings Figure 2-5 shows the E1563A logical address switch with the switch shown
in the factory setting. Set the logical address to match the VXI mainframe
slot number for a one mainframe system. For a two mainframe system, set
to the slot number plus 16, if in mainframe #1, or to the slot number plus 32,
if in mainframe #2. Refer to “VXI Module Locations And Logical
Addresses” on page 32. The Interrupt Priority is always set to 1. Both the
servant area and bus request factory defaults are correct .
Figure 2-5. E1563A Digitizer Switch and Jumper Settings