Specifications
Calibration and Preventive Maintenance 105Chapter 6
6 HVCI Gain 30.52e-6 30.0e-6 31.0e-6
7 Comp200V Gain 1.848 1.75 2.05
8 Comp20V Gain 0.1848 0.175 0.205
9 LV ClampDac Gain 78.13e-3 0.065 0.09
10 HV ClampDac Gain 430.0e-3 0.30 0.55
11 Arb MDac Gain 7.723e-3 0.005 0.009
12 Arb Xfrm Gain 38.62e-3 0.025 0.050
13 LVCV Offset -19.77 -21.0 -18.5
14 HVCV Offset -110.0 -125.0 -105.0
15 LVCI 200m Offset -988.6e-3 -1.065 -0.915
16 LVCI 20m Offset -98.86e-3 -0.104 -0.094
17 LVCI 2m Offset -9.886e-3 -0.0104 -0.0094
18 LVCI 200u Offset -988.6e-6 -0.00104 -0.00094
19 HVCI Offset -1.0 -1.05 -0.95
20 Comp200V Offset -237.5 -255.0 -220.0
21 Comp20V Offset -23.75 -24.5 -23.0
22 LV ClampDac Offset -0.7 ,0.6 2.4
23 HV ClampDac Offset -2.0 3.3 11.0
24 LV 10 10.0 10.0 12.5
25 LV 100 100.0 100.0 102.5
26 LV 1K 1000.0 990.0 1010.0
27 LV 10K 10000.0 9988.0 10018.0
28 HV 100 100.0 100.0 102.5
29 DRCI 200m 0.0 -0.05 0.45
30 DRCI 20m 0.0 -0.5 0.9
31 DRCI 2m 0.0 -5.0 5.0
32 DRCI HV20m 0.0 -0.5 0.9
33 DRFVMI 200m 0.0 -0.05 0.45
34 DRFVMI 20m 0.0 -0.5 0.9
35 DRFVMI 2m 0.0 -5.0 5.0
Table 6-2. Parameter ID Numbers and Types
ID
#
Type
Nominal
Value
Low
Value
High
Value