Specifications
100 Calibration and Preventive Maintenance Chapter 6
3. Select the OK button in the prompt box. Note, if you select the
CANCEL button, this portion of the Agilent E6171A calibration is
not performed.
4. When prompted by the ‘autoadj.tpa’ testplan , remove the test
fixture ribbon cable and reconnect the ribbon cable from the E6172A
Matrix Module.
5. Select the OK button in the prompt box.
6. The ‘autoadj.tpa’ testplan prompts with a box asking if you are
ready to adjust potentiometers R196 and 197. These two adjustments
are located on the front panel of the Agilent E6171A module. Select
OK, to proceed with the adjustments. If you select CANCEL, the
final manual adjustments to the module are not made.